    The eVolution® Ventilator is a new paradigm in ventilator design.  Dual PSOL valves ensure the precise flow, volume and pressure delivery for your sickest patients—true ICU ventilator care—while the innovative turbine design means you can deliver this ICU equivalent care wherever and whenever you need it.  The ventilator can receive O2 input from either a high or low pressure gas source, giving the eVolution precision and versatility throughout the continuum of ventilator care.

    A full suite of modes gives the clinician the flexibility to choose the mode that best suits the patient’s needs.

    The eVolution Ventilator is not available in the USA.

      eVent Medical | eVolutions® Ventilator
  •  
     

    "The night the lights went out . . . to our amazement, the Inspiration ventilators continued to ventilate our patients, uninterrupted."

    Florida

     
    Features
    • Comprehensive graphics
      and trending
    • High performance internal gas source technology
    • Dual valve technology for precise breath delivery
    • Pediatric through adult
    • Invasive and noninvasive
    • Transport ventilation support
    • Easy to use touch screen
      user interface
     

    A Clinician Focused Respiratory Product

  •  
     

    "The night the lights went out . . . to our amazement, the Inspiration ventilators continued to ventilate our patients, uninterrupted."

    Florida

     
    Gallery

eVolution Ventilator

 

 

   
 

Download
International Datasheet

  eVent Medical | eVolution® Ventilator (Int'l) Datasheet
 

Contact eVent Medical

Do you have questions about our product, would like to speak to one of our Customer Service Specialists or require additional literature?

Request Information

 

Get Product Support

Comprised mostly of respiratory therapists, our clinical and technical teams are available 24 hours a day, 7 days a week.

Get Support Now

 
 

Specifications

 
Modes of Ventilation
  • Assisted Control Mandatory Ventilation (CMV)
  • Synchronized Intermittent Mandatory Ventilation (SIMV)
  • Spontaneous Ventilation (CPAP + PS, CPAP + VS, SPAP)
 
Additional Features
  • Auto Control
  • Apnea Backup Ventilation
  • SPAP Philosophy Options: Time, Cycle + Time, Cycle + I:E
  • IBW Calculator
  • NIV On/Off in all Modes
  • Active Exhalation Valve
  • Automatic Leak Compensation (up to 60 lpm)
 
Breath Types
  • Volume-controlled
 

(V-CMV; V-SIMV)

  • Pressure-controlled
 

(P-CMV; P-SIMV, PS)

  • Volume Targeted Pressure-controlled
 

(PRVC-CMV; PRVC-SIMV, VS)

  • Dual Level PEEP
 

(SPAP)

 
Breath Triggering
  • Pressure triggering
 

-0.5 to (-20) cmH2O

  • Flow triggering
 

0.5 - 20 lpm

 
Additional Settings
  • Respiratory rate
 

0 - 120 bpm

  • Tidal volume
 

50 - 2000 ml*

  • PEEP / CPAP
 

0 - 40 cmH2O

  • Pcontrol
 

2 - 80 cmH2O*

  • Psupport
 

0 - 60 cmH2O

  • Peak flow (mandatory)
 

5 - 120 lpm*

  • Peak flow (spontaneous)
 

Up to 180 lpm*

  • I-Times
 

0.2 - 4.0 sec

  • Oxygen concentration
 

21 - 100%

  • Rise time settings
 

10 (Fast), 5 (Med), or 1 (Slow)

  • Exhalation sensitivity %
    (of spontaneous peak flow)
 

10 - 80%

  • Base Flow (with Auto leak comp off)
 

 

  • NIV On
 

2.5 - 60 lpm

  • NIV Off
 

2.5 - 25 lpm

  • Plateau (insp pause)
 

0 - 2.0 sec

 
User Configurable Graphic User Interface (GUI) and Functions
  • Multiple Easy View Tabs
    (Main, Settings, Monitoring, Alarms, Configuration)
 

 

  • Graphics Display (2-Loops and 3-Waveforms)
 

 

  • Apnea Back-up
    • User-selected breath type and settings
 

 

  • Smart Nebulizer™*
    • User-selected nebulization running time and interval
    • Volume compensation
 

 

  • Smart Sigh™
    • User-defined volume- or pressure-controlled sighs
    • User-defined frequency and multiples of sighs
    • +0 -50% of tidal volume or pressure setting
 

 

  • Humidification Selection (HME, Humidifier, and None)
 

 

 
User Initiated Functions
  • Manual inspiration
  • Alarm silence
  • 100% Oxygen
  • Inspiratory and expiratory breath holds
  • Standby
 
Monitored / Displayed Data
  • Delivered oxygen concentrations:
 

21 - 100%

  • Trend Data (1-72 hrs.) (26 trend parameters)
 

 

  • Monitoring Data (26 monitored parameters):

Basic

 

Advanced

 

Mechanics

  • Ppeak
 
  • PF
 
  • Cstat
  • PEEP
 
  • PFe
 
  • Rinsp
  • Pmean
 
  • FiO2
 
  • Rexp
  • Pplateau
 
  • H:L
 
  • Auto PEEP
  • Vte
 
  • Leak rate
 
  • RSBI
  • Vti
 
  • Ti / Ttot
   
  • Exp Min Vol (Ve)
 
  • I:E
   
  • Spont Min Vol
    (Ve Spont)
 
  • Spont % 1h
   
  • Rate Spont
 
  • Spont % 8h
   
  • Respiratory rate
       
  • Ti
       
  • Te
       

Real Time Graphics

  • Volume + time
 
  • Flow Volume loop
  • Flow + time
 
  • Pressure Volume loop
  • Pressure + time
   
 
Prioritized Alarms (High, Medium and Low)
  • Auto set alarm feature
  • Event alarm log
 
Communication Interfaces
  • Ethernet and Nurse call
 
Physical Data
  • Integrated high performance, quiet blower technology
  • AC power: 100 - 240 VAC (50/60Hz)
  • Internal and optional external battery
  • Battery backup > 120min
  • Oxygen Supply pressure 5 - 90 psi
  • Screen: LCD 12.1” / Touch Screen
 

* Patient Type Dependent

 

Specifications subject to change.

 

Support & Maintenance

Comprised mostly of respiratory therapists, our clinical and technical teams are available 24 hours a day, 7 days a week.

Support


Our Warranty

eVent Medical | 3 Year International Product Warranty  

3 Year International Product Warranty

 

Sales Inquiries

Email:
Contact Via Online Form

Telephone:
+1 949 492 8368